Evaluation of Concomitant Imaging Dose in 4D-CBCT Guided Thoracic Radiotherapy

Introduction This retrospective study evaluated the imaging dose of 4D-CBCT in patients treated with thoracic radiotherapy. Methods A geometric model of a 4D-CBCT system was built for Monte Carlo (MC) calculations. Percentage depth dose (PDD) and profiles of the system were measured and compared wit...

Full description

Saved in:
Bibliographic Details
Main Authors: Wu Jianan PhD, Sang Yong MS, Li Zhangmin MS, Liang Ying PhD, Zhao Man MS, Hu Yuchao PhD, Jia Yajun MS
Format: Article
Language:English
Published: SAGE Publishing 2025-08-01
Series:Technology in Cancer Research & Treatment
Online Access:https://doi.org/10.1177/15330338251366729
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Introduction This retrospective study evaluated the imaging dose of 4D-CBCT in patients treated with thoracic radiotherapy. Methods A geometric model of a 4D-CBCT system was built for Monte Carlo (MC) calculations. Percentage depth dose (PDD) and profiles of the system were measured and compared with MC calculations to verify the model. Eight lung cancer patients were retrospectively selected for 4D-CBCT imaging dose assessment. For each patient, the imaging dose was calculated using the CBCT log file, simulating CT and the MC model. Organ at risk (OAR) dose parameters from the imaging dose were calculated, and the effect of the imaging dose on OAR dose parameters when the imaging dose was added to the treatment dose was analyzed. Results Deviations between MC calculations and measurements for PDD and profiles were within 1.1% and 5%, respectively. For a pre-treatment 4D-CBCT scan, the mean lung and heart doses were 9.5 ± 1.4 mGy (mean value ± standard deviation, hereafter the same) and 9.7 ± 1.4 mGy, and the maximum spinal cord, esophagus and ribs doses were 17.6 ± 4.2 mGy, 17.1 ± 3.3 mGy and 37.8 ± 5.3 mGy, respectively. For an in-treatment 4D-CBCT scan, the dose parameters were 5.6 ± 1.8 mGy, 5.6 ± 1.5 mGy, 10.3 ± 3.5 mGy, 10.3 ± 4.3 mGy and 22.4 ± 5.3 mGy, respectively. The OAR dose parameters for in-treatment 4D-CBCT scans showed a relatively strong linear relationship with beam MU and number of frames. For patients receiving 5-fraction treatments and one pre-treatment and one in-treatment 4D-CBCT for each fraction, the increases in lung V20, lung V5, mean lung and heart doses, maximum spinal cord, esophagus and ribs doses were 0.04 ± 0.03%, 0.29 ± 0.23%, 76.3 ± 14.4 mGy, 76.5 ± 12.0 mGy, 76.3 ± 20.2 mGy, 70.5 ± 13.1 mGy and 101.2 ± 20.5 mGy, respectively. The increase in OAR dose parameters was proportional to the number of fractions. Conclusion The imaging dose of pre-treatment and in-treatment 4D-CBCT was calculated using a validated MC model. For 5-fraction treatments, the imaging dose will likely have minimal clinical impact on OAR dose parameters.
ISSN:1533-0338