Performance based on NEMA NU-4 2008 Standard of CDTN/CNEN’s Small Animal PET Scanner

This study aimed to evaluate the general performance of the preclinical PET scanner (LabPET Solo 4, GE) at CDTN/CNEN using the NEMA NU 4-2008 protocol. The performance parameters include: (i) spatial resolution; (ii) scatter fraction and counts rate performance; (iii) sensitivity and (iv) image qual...

Full description

Saved in:
Bibliographic Details
Main Authors: Rodrigo Modesto Gadelha Gontijo, Andréa Vidal Ferreira, Juliana Batista da Silva, Marcelo Mamede
Format: Article
Language:English
Published: Brazilian Radiation Protection Society (Sociedade Brasileira de Proteção Radiológica, SBPR) 2022-09-01
Series:Brazilian Journal of Radiation Sciences
Subjects:
Online Access:https://bjrs.org.br/revista/index.php/REVISTA/article/view/2006
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1849318478363754496
author Rodrigo Modesto Gadelha Gontijo
Andréa Vidal Ferreira
Juliana Batista da Silva
Marcelo Mamede
author_facet Rodrigo Modesto Gadelha Gontijo
Andréa Vidal Ferreira
Juliana Batista da Silva
Marcelo Mamede
author_sort Rodrigo Modesto Gadelha Gontijo
collection DOAJ
description This study aimed to evaluate the general performance of the preclinical PET scanner (LabPET Solo 4, GE) at CDTN/CNEN using the NEMA NU 4-2008 protocol. The performance parameters include: (i) spatial resolution; (ii) scatter fraction and counts rate performance; (iii) sensitivity and (iv) image quality. For the evaluation specific simulators and a sealed radioactive source of Sodium-22 were used. The results revealed that the NEMA-based radial spatial resolution ranged from 1.79 mm at the center of the FOV to 2.49 mm at a radial offset of 25mm. Tangential resolution ranged from 1.95mm at the center of the FOV to 3.50mm at a radial offset of 25mm. Axial resolution ranged from 1.78mm at the center of the FOV to 5.42mm at a radial offset of 25mm. Iterative image reconstruction improved the radial spatial resolution to 0.93mm at the center of the FOV. For the mouse-sized phantom, the peak noise equivalent count rate (NECR) was 37.8kcps at 85MBq whereas the peak true count rate was 169.9kcps at 125MBq with a count loss of 7.8% for the maximum rate reached at 103.3MBq. The peak of sensitivity (S) and absolute sensitivity (SA) at the center of axial field-of-view (FOV) is 5.7cps.Bq-1 and 0.63%, respectively. The overall imaging capabilities of the scanner were assessed using the NEMA image-quality phantom. The image roughness was 9.5%. Spillover rates (SORs) in air and in water were 0.26 and 0.17 respectively. The recovery coefficients (RCs) were 0.12 for the 1-mm-diameter rod and 0.91 for the 5-mm-diameter rod. The results demonstrate that CDTN/CNENsmall animal PET scanner has adequate reliability and acquired images are compatible with international quality standards for the molecular imaging research.
format Article
id doaj-art-419ec22d3de7436aa2a6be3c778ee9a4
institution Kabale University
issn 2319-0612
language English
publishDate 2022-09-01
publisher Brazilian Radiation Protection Society (Sociedade Brasileira de Proteção Radiológica, SBPR)
record_format Article
series Brazilian Journal of Radiation Sciences
spelling doaj-art-419ec22d3de7436aa2a6be3c778ee9a42025-08-20T03:50:49ZengBrazilian Radiation Protection Society (Sociedade Brasileira de Proteção Radiológica, SBPR)Brazilian Journal of Radiation Sciences2319-06122022-09-0110310.15392/bjrs.v10i3.20061624Performance based on NEMA NU-4 2008 Standard of CDTN/CNEN’s Small Animal PET ScannerRodrigo Modesto Gadelha Gontijo0Andréa Vidal Ferreira1Juliana Batista da Silva2Marcelo Mamede3Federal University of Minas GeraisCentro de Desenvolvimento da Tecnologia Nuclear (CDTN)Centro de Desenvolvimento da Tecnologia Nuclear (CDTN)UFMGThis study aimed to evaluate the general performance of the preclinical PET scanner (LabPET Solo 4, GE) at CDTN/CNEN using the NEMA NU 4-2008 protocol. The performance parameters include: (i) spatial resolution; (ii) scatter fraction and counts rate performance; (iii) sensitivity and (iv) image quality. For the evaluation specific simulators and a sealed radioactive source of Sodium-22 were used. The results revealed that the NEMA-based radial spatial resolution ranged from 1.79 mm at the center of the FOV to 2.49 mm at a radial offset of 25mm. Tangential resolution ranged from 1.95mm at the center of the FOV to 3.50mm at a radial offset of 25mm. Axial resolution ranged from 1.78mm at the center of the FOV to 5.42mm at a radial offset of 25mm. Iterative image reconstruction improved the radial spatial resolution to 0.93mm at the center of the FOV. For the mouse-sized phantom, the peak noise equivalent count rate (NECR) was 37.8kcps at 85MBq whereas the peak true count rate was 169.9kcps at 125MBq with a count loss of 7.8% for the maximum rate reached at 103.3MBq. The peak of sensitivity (S) and absolute sensitivity (SA) at the center of axial field-of-view (FOV) is 5.7cps.Bq-1 and 0.63%, respectively. The overall imaging capabilities of the scanner were assessed using the NEMA image-quality phantom. The image roughness was 9.5%. Spillover rates (SORs) in air and in water were 0.26 and 0.17 respectively. The recovery coefficients (RCs) were 0.12 for the 1-mm-diameter rod and 0.91 for the 5-mm-diameter rod. The results demonstrate that CDTN/CNENsmall animal PET scanner has adequate reliability and acquired images are compatible with international quality standards for the molecular imaging research.https://bjrs.org.br/revista/index.php/REVISTA/article/view/2006quality controlnema nu 4-2008preclinical pet
spellingShingle Rodrigo Modesto Gadelha Gontijo
Andréa Vidal Ferreira
Juliana Batista da Silva
Marcelo Mamede
Performance based on NEMA NU-4 2008 Standard of CDTN/CNEN’s Small Animal PET Scanner
Brazilian Journal of Radiation Sciences
quality control
nema nu 4-2008
preclinical pet
title Performance based on NEMA NU-4 2008 Standard of CDTN/CNEN’s Small Animal PET Scanner
title_full Performance based on NEMA NU-4 2008 Standard of CDTN/CNEN’s Small Animal PET Scanner
title_fullStr Performance based on NEMA NU-4 2008 Standard of CDTN/CNEN’s Small Animal PET Scanner
title_full_unstemmed Performance based on NEMA NU-4 2008 Standard of CDTN/CNEN’s Small Animal PET Scanner
title_short Performance based on NEMA NU-4 2008 Standard of CDTN/CNEN’s Small Animal PET Scanner
title_sort performance based on nema nu 4 2008 standard of cdtn cnen s small animal pet scanner
topic quality control
nema nu 4-2008
preclinical pet
url https://bjrs.org.br/revista/index.php/REVISTA/article/view/2006
work_keys_str_mv AT rodrigomodestogadelhagontijo performancebasedonnemanu42008standardofcdtncnenssmallanimalpetscanner
AT andreavidalferreira performancebasedonnemanu42008standardofcdtncnenssmallanimalpetscanner
AT julianabatistadasilva performancebasedonnemanu42008standardofcdtncnenssmallanimalpetscanner
AT marcelomamede performancebasedonnemanu42008standardofcdtncnenssmallanimalpetscanner