The SwiftScan step-and-shoot continuous mode improves SPECT scanning efficiency: a preliminary phantom and clinical test

Abstract Purpose The aim of the study was to investigate the value of SwiftScan Step-and-Shoot Continuous (SSC) scanning mode in enhancing image quality and to explore appropriate scanning parameters for reducing scan time. Methods This study was composed of a phantom study and two clinical tests. T...

Full description

Saved in:
Bibliographic Details
Main Authors: Jicheng Li, Kai Zhang, Xingru Pang, Lele Huang, Xiaoxue Tian, Jiangyan Liu
Format: Article
Language:English
Published: SpringerOpen 2025-01-01
Series:EJNMMI Physics
Subjects:
Online Access:https://doi.org/10.1186/s40658-024-00709-0
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1841559055538061312
author Jicheng Li
Kai Zhang
Xingru Pang
Lele Huang
Xiaoxue Tian
Jiangyan Liu
author_facet Jicheng Li
Kai Zhang
Xingru Pang
Lele Huang
Xiaoxue Tian
Jiangyan Liu
author_sort Jicheng Li
collection DOAJ
description Abstract Purpose The aim of the study was to investigate the value of SwiftScan Step-and-Shoot Continuous (SSC) scanning mode in enhancing image quality and to explore appropriate scanning parameters for reducing scan time. Methods This study was composed of a phantom study and two clinical tests. The differences in visual image quality scores, coefficient of variance (COV) of the background, image signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and recovery coefficient (RC) of the sphere were compared between SSC mode and traditional Step-and-Shoot (SS) mode in the phantom study. Various “shoot” acquisition times (5s, 10s, 15s) and “step” angles (3-degree, 6-degree, 9-degree) were evaluated and verified. In the clinical tests, bone tomography and parathyroid tomography were performed on 30 patients each. Differences in visual image quality scores, background COV, image SNR, CNR, and standardized uptake value (SUV) of lesions were compared between the two modes. Results In the phantom study, SSC mode demonstrated higher visual scores and significantly reduced background COV (P < 0.05), and significantly increased SNR and CNR (P < 0.05) compared to SS mode. No significant alteration in RC was observed (P > 0.05). In the clinical tests, no significant differences were found between the optimal SSC scan combination (10s “shoot” and 6-degree “step”)/ (10s “shoot” and 3-degree “step”) and the traditional SS scan combination (15s “shoot” and 6-degree “step”)/ (15s “shoot” and 3-degree “step”) in visual image quality scores, background COV, image SNR, CNR, and SUV of bone and parathyroid high uptake lesions (P > 0.05). Conclusion The SwiftScan SSC mode can reduce acquisition time by 33% while maintaining similar image quality and quantification accuracy compared to SS mode. An SSC scanning protocol with a 10s “shoot” acquisition and 6-degree “step” or with a 10s “shoot” acquisition and 3-degree “step” over a 360-degree rotation, is recommended for clinical use.
format Article
id doaj-art-3f79ccda7e34420cbc5a4b737eb74c06
institution Kabale University
issn 2197-7364
language English
publishDate 2025-01-01
publisher SpringerOpen
record_format Article
series EJNMMI Physics
spelling doaj-art-3f79ccda7e34420cbc5a4b737eb74c062025-01-05T12:46:39ZengSpringerOpenEJNMMI Physics2197-73642025-01-0112111810.1186/s40658-024-00709-0The SwiftScan step-and-shoot continuous mode improves SPECT scanning efficiency: a preliminary phantom and clinical testJicheng Li0Kai Zhang1Xingru Pang2Lele Huang3Xiaoxue Tian4Jiangyan Liu5Department of Nuclear Medicine, Lanzhou University Second HospitalDepartment of Nuclear Medicine, Lanzhou University Second HospitalDepartment of Nuclear Medicine, Lanzhou University Second HospitalDepartment of Nuclear Medicine, Lanzhou University Second HospitalDepartment of Nuclear Medicine, Lanzhou University Second HospitalDepartment of Nuclear Medicine, Lanzhou University Second HospitalAbstract Purpose The aim of the study was to investigate the value of SwiftScan Step-and-Shoot Continuous (SSC) scanning mode in enhancing image quality and to explore appropriate scanning parameters for reducing scan time. Methods This study was composed of a phantom study and two clinical tests. The differences in visual image quality scores, coefficient of variance (COV) of the background, image signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and recovery coefficient (RC) of the sphere were compared between SSC mode and traditional Step-and-Shoot (SS) mode in the phantom study. Various “shoot” acquisition times (5s, 10s, 15s) and “step” angles (3-degree, 6-degree, 9-degree) were evaluated and verified. In the clinical tests, bone tomography and parathyroid tomography were performed on 30 patients each. Differences in visual image quality scores, background COV, image SNR, CNR, and standardized uptake value (SUV) of lesions were compared between the two modes. Results In the phantom study, SSC mode demonstrated higher visual scores and significantly reduced background COV (P < 0.05), and significantly increased SNR and CNR (P < 0.05) compared to SS mode. No significant alteration in RC was observed (P > 0.05). In the clinical tests, no significant differences were found between the optimal SSC scan combination (10s “shoot” and 6-degree “step”)/ (10s “shoot” and 3-degree “step”) and the traditional SS scan combination (15s “shoot” and 6-degree “step”)/ (15s “shoot” and 3-degree “step”) in visual image quality scores, background COV, image SNR, CNR, and SUV of bone and parathyroid high uptake lesions (P > 0.05). Conclusion The SwiftScan SSC mode can reduce acquisition time by 33% while maintaining similar image quality and quantification accuracy compared to SS mode. An SSC scanning protocol with a 10s “shoot” acquisition and 6-degree “step” or with a 10s “shoot” acquisition and 3-degree “step” over a 360-degree rotation, is recommended for clinical use.https://doi.org/10.1186/s40658-024-00709-0SwiftScanStep-and-shoot continuous modeSPECT/CTImage qualityAcquisition efficiency
spellingShingle Jicheng Li
Kai Zhang
Xingru Pang
Lele Huang
Xiaoxue Tian
Jiangyan Liu
The SwiftScan step-and-shoot continuous mode improves SPECT scanning efficiency: a preliminary phantom and clinical test
EJNMMI Physics
SwiftScan
Step-and-shoot continuous mode
SPECT/CT
Image quality
Acquisition efficiency
title The SwiftScan step-and-shoot continuous mode improves SPECT scanning efficiency: a preliminary phantom and clinical test
title_full The SwiftScan step-and-shoot continuous mode improves SPECT scanning efficiency: a preliminary phantom and clinical test
title_fullStr The SwiftScan step-and-shoot continuous mode improves SPECT scanning efficiency: a preliminary phantom and clinical test
title_full_unstemmed The SwiftScan step-and-shoot continuous mode improves SPECT scanning efficiency: a preliminary phantom and clinical test
title_short The SwiftScan step-and-shoot continuous mode improves SPECT scanning efficiency: a preliminary phantom and clinical test
title_sort swiftscan step and shoot continuous mode improves spect scanning efficiency a preliminary phantom and clinical test
topic SwiftScan
Step-and-shoot continuous mode
SPECT/CT
Image quality
Acquisition efficiency
url https://doi.org/10.1186/s40658-024-00709-0
work_keys_str_mv AT jichengli theswiftscanstepandshootcontinuousmodeimprovesspectscanningefficiencyapreliminaryphantomandclinicaltest
AT kaizhang theswiftscanstepandshootcontinuousmodeimprovesspectscanningefficiencyapreliminaryphantomandclinicaltest
AT xingrupang theswiftscanstepandshootcontinuousmodeimprovesspectscanningefficiencyapreliminaryphantomandclinicaltest
AT lelehuang theswiftscanstepandshootcontinuousmodeimprovesspectscanningefficiencyapreliminaryphantomandclinicaltest
AT xiaoxuetian theswiftscanstepandshootcontinuousmodeimprovesspectscanningefficiencyapreliminaryphantomandclinicaltest
AT jiangyanliu theswiftscanstepandshootcontinuousmodeimprovesspectscanningefficiencyapreliminaryphantomandclinicaltest
AT jichengli swiftscanstepandshootcontinuousmodeimprovesspectscanningefficiencyapreliminaryphantomandclinicaltest
AT kaizhang swiftscanstepandshootcontinuousmodeimprovesspectscanningefficiencyapreliminaryphantomandclinicaltest
AT xingrupang swiftscanstepandshootcontinuousmodeimprovesspectscanningefficiencyapreliminaryphantomandclinicaltest
AT lelehuang swiftscanstepandshootcontinuousmodeimprovesspectscanningefficiencyapreliminaryphantomandclinicaltest
AT xiaoxuetian swiftscanstepandshootcontinuousmodeimprovesspectscanningefficiencyapreliminaryphantomandclinicaltest
AT jiangyanliu swiftscanstepandshootcontinuousmodeimprovesspectscanningefficiencyapreliminaryphantomandclinicaltest