COMPENSATION OF MEASUREMENT ERRORS WHEN REDUCING LINEAR DIMENSIONS OF THE KELVIN PROBE
The study is based on results of modeling of measurement circuit containing vibrating-plate capacitor using a complex-harmonic analysis technique. Low value of normalized frequency of small-sized scanning Kelvin probe leads to high distortion factor of probe’s measurement signal that in turn leads t...
Saved in:
| Main Authors: | A. K. Tyavlovsky, A. L. Zharin |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
Belarusian National Technical University
2015-03-01
|
| Series: | Приборы и методы измерений |
| Subjects: | |
| Online Access: | https://pimi.bntu.by/jour/article/view/36 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Similar Items
-
KELVIN PROBE’S STRAY CAPACITANCE AND NOISE SIMULATION
by: S. Danyluk, et al.
Published: (2015-03-01) -
ANALYSIS OF SURFACE DEFECTS OF ALUMINUM AND ITS ALLOYS WITH A SCANNING KELVIN PROBE
by: A. K. Tyavlovsky, et al.
Published: (2017-02-01) -
MATHEMATICAL MODELING OF A DISTANCE DEPENDENCE OF A SCANNING KELVIN PROBE LATERAL RESOLUTION
by: A. K. Tyavlovsky
Published: (2015-04-01) -
Surface Electric Potential Measurement with a Static Probe
by: R. I. Vorobey, et al.
Published: (2023-08-01) -
Detection of Hidden Defects Induced by Thermomechanical Processing of Aluminum Substrates for Sensor Devices Using a Scanning Kelvin Probe
by: A. L Zharin, et al.
Published: (2025-03-01)