Evaluation of a Digital Mammography System with the MAM-162 phantom

Quality controls are performed to verify the performance status of X-ray equipment in relation to image quality and patient dose. Originally, the Laboratory of Radiation Physics and Metrology (LAF-RAM) in Nicaragua had only one phantom to evaluate mammography image quality subjectively; however, und...

Full description

Saved in:
Bibliographic Details
Main Authors: A. Castillo, N. A. Roas Zúniga
Format: Article
Language:English
Published: Brazilian Radiation Protection Society (Sociedade Brasileira de Proteção Radiológica, SBPR) 2024-02-01
Series:Brazilian Journal of Radiation Sciences
Subjects:
Online Access:https://bjrs.org.br/revista/index.php/REVISTA/article/view/2283
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1849423870757437440
author A. Castillo
N. A. Roas Zúniga
author_facet A. Castillo
N. A. Roas Zúniga
author_sort A. Castillo
collection DOAJ
description Quality controls are performed to verify the performance status of X-ray equipment in relation to image quality and patient dose. Originally, the Laboratory of Radiation Physics and Metrology (LAF-RAM) in Nicaragua had only one phantom to evaluate mammography image quality subjectively; however, under a cooperation project with the International Atomic Energy Agency (IAEA), the MAM-162 phantom was acquired for the evaluation of physical image quality. In this work a CR type digital mammography system was evaluated using the MAM-162 phantom with adaptation to a single quality control protocol based on international standards such as the German Institute for Standardization (DIN) DIN 6868-162:2013-06, the IAEA technical document TECDOC protocol No. 1958 and the European Federation of Organizations for Medical Physics (EFOMP) mammo protocol. Among the main tests performed are the evaluation of the lateral image limit of the chest wall with satisfactory results, the automatic exposure control, the Signal-to-Noise difference ratio that complies with the tolerances, in the case Ghosting and dynamic range the results of both are satisfactory, the presence of artifacts was observed when applying variance mapping, a base value of the dynamic range was also obtained. For the dosimetry, the half value layer and the determination of the Mean Glandular Dose were evaluated, which does not comply with the tolerances. The final protocol document could be validated and implemented as part of the management and service documents of the Quality Control and Monitoring Laboratory (LCM) of LAF-RAM.
format Article
id doaj-art-3343ff590fef4b839bc5100522113ca2
institution Kabale University
issn 2319-0612
language English
publishDate 2024-02-01
publisher Brazilian Radiation Protection Society (Sociedade Brasileira de Proteção Radiológica, SBPR)
record_format Article
series Brazilian Journal of Radiation Sciences
spelling doaj-art-3343ff590fef4b839bc5100522113ca22025-08-20T03:30:25ZengBrazilian Radiation Protection Society (Sociedade Brasileira de Proteção Radiológica, SBPR)Brazilian Journal of Radiation Sciences2319-06122024-02-01121A (Suppl.)e2283e228310.15392/2319-0612.2024.22831905Evaluation of a Digital Mammography System with the MAM-162 phantomA. Castillo0N. A. Roas Zúniga1Universidad Nacional Autónoma de Nicaragua, ManaguaUniversidad Nacional Autónoma de Nicaragua, ManaguaQuality controls are performed to verify the performance status of X-ray equipment in relation to image quality and patient dose. Originally, the Laboratory of Radiation Physics and Metrology (LAF-RAM) in Nicaragua had only one phantom to evaluate mammography image quality subjectively; however, under a cooperation project with the International Atomic Energy Agency (IAEA), the MAM-162 phantom was acquired for the evaluation of physical image quality. In this work a CR type digital mammography system was evaluated using the MAM-162 phantom with adaptation to a single quality control protocol based on international standards such as the German Institute for Standardization (DIN) DIN 6868-162:2013-06, the IAEA technical document TECDOC protocol No. 1958 and the European Federation of Organizations for Medical Physics (EFOMP) mammo protocol. Among the main tests performed are the evaluation of the lateral image limit of the chest wall with satisfactory results, the automatic exposure control, the Signal-to-Noise difference ratio that complies with the tolerances, in the case Ghosting and dynamic range the results of both are satisfactory, the presence of artifacts was observed when applying variance mapping, a base value of the dynamic range was also obtained. For the dosimetry, the half value layer and the determination of the Mean Glandular Dose were evaluated, which does not comply with the tolerances. The final protocol document could be validated and implemented as part of the management and service documents of the Quality Control and Monitoring Laboratory (LCM) of LAF-RAM.https://bjrs.org.br/revista/index.php/REVISTA/article/view/2283din 6868-162:2013-06quality controlmam-162
spellingShingle A. Castillo
N. A. Roas Zúniga
Evaluation of a Digital Mammography System with the MAM-162 phantom
Brazilian Journal of Radiation Sciences
din 6868-162:2013-06
quality control
mam-162
title Evaluation of a Digital Mammography System with the MAM-162 phantom
title_full Evaluation of a Digital Mammography System with the MAM-162 phantom
title_fullStr Evaluation of a Digital Mammography System with the MAM-162 phantom
title_full_unstemmed Evaluation of a Digital Mammography System with the MAM-162 phantom
title_short Evaluation of a Digital Mammography System with the MAM-162 phantom
title_sort evaluation of a digital mammography system with the mam 162 phantom
topic din 6868-162:2013-06
quality control
mam-162
url https://bjrs.org.br/revista/index.php/REVISTA/article/view/2283
work_keys_str_mv AT acastillo evaluationofadigitalmammographysystemwiththemam162phantom
AT naroaszuniga evaluationofadigitalmammographysystemwiththemam162phantom