Low background facility for ionizing radiation at LNE LNHB: Design and traceability
This paper presents the set up of a low background calibration facility aimed at calibrating environment survey meter. The facility is a lead box 100 cm x 100 cm x 125 cm with 2.5 cm thickness walls, the Caesium 137 source is located in a conical collimator including an internal cavity for the sourc...
Saved in:
| Main Authors: | , , , , |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
EDP Sciences
2025-01-01
|
| Series: | EPJ Web of Conferences |
| Online Access: | https://www.epj-conferences.org/articles/epjconf/pdf/2025/08/epjconf_cim2025_11001.pdf |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| Summary: | This paper presents the set up of a low background calibration facility aimed at calibrating environment survey meter. The facility is a lead box 100 cm x 100 cm x 125 cm with 2.5 cm thickness walls, the Caesium 137 source is located in a conical collimator including an internal cavity for the source. The ambient dose equivalent rate in the beam of this facility varies between 300 and 80 nSv/h. This quantity is traceable to the air kerma national references of LNE-LNHB. The standard uncertainty for the ambient dose equivalent rate is about 2% and the uncertainty of the calibration coefficient od a survey meter, depending meanly on the uncertainty if the background measure, is estimated between 5% and 10%. |
|---|---|
| ISSN: | 2100-014X |