Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
Failure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms...
Saved in:
| Main Authors: | , , , |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
Medycyna Praktyczna
2016-04-01
|
| Series: | Videosurgery and Other Miniinvasive Techniques |
| Subjects: | |
| Online Access: | https://www.termedia.pl/Acute-limb-ischemia-caused-by-incorrect-deployment-of-a-clip-based-arterial-closure-device,42,27446,1,1.html |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| _version_ | 1850230708065271808 |
|---|---|
| author | Łukasz Dzieciuchowicz Maciej Zmysłowski Karolina Stefaniak Grzegorz Oszkinis |
| author_facet | Łukasz Dzieciuchowicz Maciej Zmysłowski Karolina Stefaniak Grzegorz Oszkinis |
| author_sort | Łukasz Dzieciuchowicz |
| collection | DOAJ |
| description | Failure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms of acute limb ischemia occurred at the start of the ambulation, 6 h after completion of the procedure. Because of the severity of ischemia the patient was treated surgically, and limb perfusion was successfully restored. An attempt of closure of an inadvertently punctured narrow superficial femoral artery was identified as the cause of this complication. |
| format | Article |
| id | doaj-art-1e50f1c12a0042ee8dab5f75babc5bff |
| institution | OA Journals |
| issn | 1895-4588 2299-0054 |
| language | English |
| publishDate | 2016-04-01 |
| publisher | Medycyna Praktyczna |
| record_format | Article |
| series | Videosurgery and Other Miniinvasive Techniques |
| spelling | doaj-art-1e50f1c12a0042ee8dab5f75babc5bff2025-08-20T02:03:47ZengMedycyna PraktycznaVideosurgery and Other Miniinvasive Techniques1895-45882299-00542016-04-0111211111410.5114/wiitm.2016.5957727446Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure deviceŁukasz DzieciuchowiczMaciej ZmysłowskiKarolina StefaniakGrzegorz OszkinisFailure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms of acute limb ischemia occurred at the start of the ambulation, 6 h after completion of the procedure. Because of the severity of ischemia the patient was treated surgically, and limb perfusion was successfully restored. An attempt of closure of an inadvertently punctured narrow superficial femoral artery was identified as the cause of this complication.https://www.termedia.pl/Acute-limb-ischemia-caused-by-incorrect-deployment-of-a-clip-based-arterial-closure-device,42,27446,1,1.html<i>vascular injury endovascular treatment femoral artery injuries peripheral arterial catheterization</i> |
| spellingShingle | Łukasz Dzieciuchowicz Maciej Zmysłowski Karolina Stefaniak Grzegorz Oszkinis Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device Videosurgery and Other Miniinvasive Techniques <i>vascular injury endovascular treatment femoral artery injuries peripheral arterial catheterization</i> |
| title | Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device |
| title_full | Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device |
| title_fullStr | Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device |
| title_full_unstemmed | Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device |
| title_short | Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device |
| title_sort | acute limb ischemia caused by incorrect deployment of a clip based arterial closure device |
| topic | <i>vascular injury endovascular treatment femoral artery injuries peripheral arterial catheterization</i> |
| url | https://www.termedia.pl/Acute-limb-ischemia-caused-by-incorrect-deployment-of-a-clip-based-arterial-closure-device,42,27446,1,1.html |
| work_keys_str_mv | AT łukaszdzieciuchowicz acutelimbischemiacausedbyincorrectdeploymentofaclipbasedarterialclosuredevice AT maciejzmysłowski acutelimbischemiacausedbyincorrectdeploymentofaclipbasedarterialclosuredevice AT karolinastefaniak acutelimbischemiacausedbyincorrectdeploymentofaclipbasedarterialclosuredevice AT grzegorzoszkinis acutelimbischemiacausedbyincorrectdeploymentofaclipbasedarterialclosuredevice |