Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device

Failure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms...

Full description

Saved in:
Bibliographic Details
Main Authors: Łukasz Dzieciuchowicz, Maciej Zmysłowski, Karolina Stefaniak, Grzegorz Oszkinis
Format: Article
Language:English
Published: Medycyna Praktyczna 2016-04-01
Series:Videosurgery and Other Miniinvasive Techniques
Subjects:
Online Access:https://www.termedia.pl/Acute-limb-ischemia-caused-by-incorrect-deployment-of-a-clip-based-arterial-closure-device,42,27446,1,1.html
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850230708065271808
author Łukasz Dzieciuchowicz
Maciej Zmysłowski
Karolina Stefaniak
Grzegorz Oszkinis
author_facet Łukasz Dzieciuchowicz
Maciej Zmysłowski
Karolina Stefaniak
Grzegorz Oszkinis
author_sort Łukasz Dzieciuchowicz
collection DOAJ
description Failure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms of acute limb ischemia occurred at the start of the ambulation, 6 h after completion of the procedure. Because of the severity of ischemia the patient was treated surgically, and limb perfusion was successfully restored. An attempt of closure of an inadvertently punctured narrow superficial femoral artery was identified as the cause of this complication.
format Article
id doaj-art-1e50f1c12a0042ee8dab5f75babc5bff
institution OA Journals
issn 1895-4588
2299-0054
language English
publishDate 2016-04-01
publisher Medycyna Praktyczna
record_format Article
series Videosurgery and Other Miniinvasive Techniques
spelling doaj-art-1e50f1c12a0042ee8dab5f75babc5bff2025-08-20T02:03:47ZengMedycyna PraktycznaVideosurgery and Other Miniinvasive Techniques1895-45882299-00542016-04-0111211111410.5114/wiitm.2016.5957727446Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure deviceŁukasz DzieciuchowiczMaciej ZmysłowskiKarolina StefaniakGrzegorz OszkinisFailure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms of acute limb ischemia occurred at the start of the ambulation, 6 h after completion of the procedure. Because of the severity of ischemia the patient was treated surgically, and limb perfusion was successfully restored. An attempt of closure of an inadvertently punctured narrow superficial femoral artery was identified as the cause of this complication.https://www.termedia.pl/Acute-limb-ischemia-caused-by-incorrect-deployment-of-a-clip-based-arterial-closure-device,42,27446,1,1.html<i>vascular injury endovascular treatment femoral artery injuries peripheral arterial catheterization</i>
spellingShingle Łukasz Dzieciuchowicz
Maciej Zmysłowski
Karolina Stefaniak
Grzegorz Oszkinis
Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
Videosurgery and Other Miniinvasive Techniques
<i>vascular injury
endovascular treatment
femoral artery injuries
peripheral arterial catheterization</i>
title Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
title_full Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
title_fullStr Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
title_full_unstemmed Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
title_short Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
title_sort acute limb ischemia caused by incorrect deployment of a clip based arterial closure device
topic <i>vascular injury
endovascular treatment
femoral artery injuries
peripheral arterial catheterization</i>
url https://www.termedia.pl/Acute-limb-ischemia-caused-by-incorrect-deployment-of-a-clip-based-arterial-closure-device,42,27446,1,1.html
work_keys_str_mv AT łukaszdzieciuchowicz acutelimbischemiacausedbyincorrectdeploymentofaclipbasedarterialclosuredevice
AT maciejzmysłowski acutelimbischemiacausedbyincorrectdeploymentofaclipbasedarterialclosuredevice
AT karolinastefaniak acutelimbischemiacausedbyincorrectdeploymentofaclipbasedarterialclosuredevice
AT grzegorzoszkinis acutelimbischemiacausedbyincorrectdeploymentofaclipbasedarterialclosuredevice