Twisting of Inflatable Penile Prosthesis Tubing Leading to Device Malfunction and Required Explantation: A Rare Complication

Erectile dysfunction (ED), the impairment of achieving and maintaining an erection for satisfactory sexual intercourse, is a common pathology that men experience for a variety of different factors. Conservative treatment for ED includes changing medications, lifestyle modifications, and psychotherap...

Full description

Saved in:
Bibliographic Details
Main Authors: Jordan Sarver, Eriel Emmer, Alex Benben, Matthew Skalak, Daniel Talley, Mazen Abdelhady
Format: Article
Language:English
Published: Wiley 2024-01-01
Series:Case Reports in Urology
Online Access:http://dx.doi.org/10.1155/2024/4446878
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850237650433212416
author Jordan Sarver
Eriel Emmer
Alex Benben
Matthew Skalak
Daniel Talley
Mazen Abdelhady
author_facet Jordan Sarver
Eriel Emmer
Alex Benben
Matthew Skalak
Daniel Talley
Mazen Abdelhady
author_sort Jordan Sarver
collection DOAJ
description Erectile dysfunction (ED), the impairment of achieving and maintaining an erection for satisfactory sexual intercourse, is a common pathology that men experience for a variety of different factors. Conservative treatment for ED includes changing medications, lifestyle modifications, and psychotherapy. Pharmaceutical and nonsurgical interventions include phosphodiesterase-5 inhibitors(PDE-5i), intracavernosal medication injections, and vacuum devices. Surgical treatment options for ED have evolved over time and currently include the use of inflatable penile prosthesis (IPP) and malleable penile prosthesis. IPP insertion is usually met with good patient satisfaction. However, complications of device insertion can include corporal perforation, urethral injury, cylinder erosion or extrusion, infection, and mechanical failure, to name a few. Our patient presented with device malfunction and intraoperative assessment showed the IPP tubing twisted at the levels of the reservoir on the first operation and the level of the cylinder and scrotal pump on the second operation. The twisting of the tubing resulted in a nonfunctioning IPP as the fluid was unable to fill the cylinders resulting in an erection. The patient was managed with complete device explanation and reinsertion of a new three-piece IPP per the patient and partner’s request. This is the first case report highlighting this specific complication, and we hope to provide clinicians with the resources to recognize this rare complication.
format Article
id doaj-art-1ce880b64f4f41709eaed5157744f4b4
institution OA Journals
issn 2090-6978
language English
publishDate 2024-01-01
publisher Wiley
record_format Article
series Case Reports in Urology
spelling doaj-art-1ce880b64f4f41709eaed5157744f4b42025-08-20T02:01:42ZengWileyCase Reports in Urology2090-69782024-01-01202410.1155/2024/4446878Twisting of Inflatable Penile Prosthesis Tubing Leading to Device Malfunction and Required Explantation: A Rare ComplicationJordan Sarver0Eriel Emmer1Alex Benben2Matthew Skalak3Daniel Talley4Mazen Abdelhady5Urology ResidencyMichigan State College of Osteopathic MedicineUrology ResidencyUrology ResidencyUrology ResidencyUrology ResidencyErectile dysfunction (ED), the impairment of achieving and maintaining an erection for satisfactory sexual intercourse, is a common pathology that men experience for a variety of different factors. Conservative treatment for ED includes changing medications, lifestyle modifications, and psychotherapy. Pharmaceutical and nonsurgical interventions include phosphodiesterase-5 inhibitors(PDE-5i), intracavernosal medication injections, and vacuum devices. Surgical treatment options for ED have evolved over time and currently include the use of inflatable penile prosthesis (IPP) and malleable penile prosthesis. IPP insertion is usually met with good patient satisfaction. However, complications of device insertion can include corporal perforation, urethral injury, cylinder erosion or extrusion, infection, and mechanical failure, to name a few. Our patient presented with device malfunction and intraoperative assessment showed the IPP tubing twisted at the levels of the reservoir on the first operation and the level of the cylinder and scrotal pump on the second operation. The twisting of the tubing resulted in a nonfunctioning IPP as the fluid was unable to fill the cylinders resulting in an erection. The patient was managed with complete device explanation and reinsertion of a new three-piece IPP per the patient and partner’s request. This is the first case report highlighting this specific complication, and we hope to provide clinicians with the resources to recognize this rare complication.http://dx.doi.org/10.1155/2024/4446878
spellingShingle Jordan Sarver
Eriel Emmer
Alex Benben
Matthew Skalak
Daniel Talley
Mazen Abdelhady
Twisting of Inflatable Penile Prosthesis Tubing Leading to Device Malfunction and Required Explantation: A Rare Complication
Case Reports in Urology
title Twisting of Inflatable Penile Prosthesis Tubing Leading to Device Malfunction and Required Explantation: A Rare Complication
title_full Twisting of Inflatable Penile Prosthesis Tubing Leading to Device Malfunction and Required Explantation: A Rare Complication
title_fullStr Twisting of Inflatable Penile Prosthesis Tubing Leading to Device Malfunction and Required Explantation: A Rare Complication
title_full_unstemmed Twisting of Inflatable Penile Prosthesis Tubing Leading to Device Malfunction and Required Explantation: A Rare Complication
title_short Twisting of Inflatable Penile Prosthesis Tubing Leading to Device Malfunction and Required Explantation: A Rare Complication
title_sort twisting of inflatable penile prosthesis tubing leading to device malfunction and required explantation a rare complication
url http://dx.doi.org/10.1155/2024/4446878
work_keys_str_mv AT jordansarver twistingofinflatablepenileprosthesistubingleadingtodevicemalfunctionandrequiredexplantationararecomplication
AT erielemmer twistingofinflatablepenileprosthesistubingleadingtodevicemalfunctionandrequiredexplantationararecomplication
AT alexbenben twistingofinflatablepenileprosthesistubingleadingtodevicemalfunctionandrequiredexplantationararecomplication
AT matthewskalak twistingofinflatablepenileprosthesistubingleadingtodevicemalfunctionandrequiredexplantationararecomplication
AT danieltalley twistingofinflatablepenileprosthesistubingleadingtodevicemalfunctionandrequiredexplantationararecomplication
AT mazenabdelhady twistingofinflatablepenileprosthesistubingleadingtodevicemalfunctionandrequiredexplantationararecomplication