Plasma/Serum Electrolyte and Metabolite Testing on Blood Gas Analyzer ABL837, a New Application
<b>Background</b>: Core laboratory chemistry analyzers typically use plasma and serum samples, while blood gas instruments use whole blood for electrolyte and metabolite tests. Due to high costs to back up the core lab chemistry analyzers, especially in the remote small community hospita...
Saved in:
| Main Authors: | , , |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
MDPI AG
2025-07-01
|
| Series: | Diagnostics |
| Subjects: | |
| Online Access: | https://www.mdpi.com/2075-4418/15/15/1923 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| Summary: | <b>Background</b>: Core laboratory chemistry analyzers typically use plasma and serum samples, while blood gas instruments use whole blood for electrolyte and metabolite tests. Due to high costs to back up the core lab chemistry analyzers, especially in the remote small community hospitals, we have verified the interchangeability of serum/plasma electrolytes and metabolites on blood gas instruments (GEM4000 and Radiometer ABL90) vs. chemistry analyzers. In this study, we sought to extend the investigation to another blood gas device—Radiometer ABL837. <b>Methods</b>: One plasma separator tube and one serum separator tube were drawn from 20 apparently healthy individuals and outpatients and 20 intensive care unit patients. All the samples were run on Roche Cobas8000, and then were run on three Radiometer ABL837 analyzers for sodium (Na<sup>+</sup>), potassium (K<sup>+</sup>), chloride (Cl<sup>−</sup>), glucose, lactate (plasma only), and creatinine parameters. Paired measurements between the ABL837 and Cobas8000 were compared, and their difference were assessed for statistical and clinical significance. <b>Results</b>: ABL837 demonstrated statistical significance (<i>p</i> < 0.05) vs. Cobas8000 on all the plasma and serum parameters. However, no parameter differences were found when comparing the plasma/serum results on ABL837 to those on Cobas8000, indicating that none were clinically significant. ABL837 also demonstrated good–excellent correlations with Cobas8000 on all the parameters. <b>Conclusions</b>: When comparing metabolite and electrolyte values with plasma and serum sample types, the ABL837 blood gas instruments and Cobas 8000 chemistry analyzer are interchangeable. These data proves that ABL837 can be used as a backup for a chemistry analyzer in measuring plasma and serum electrolyte and metabolite concentrations. |
|---|---|
| ISSN: | 2075-4418 |