Realistic Monte Carlo simulation of the INOR Therapy Unit and electron contamination calculation for a board 60Co Beam

The therapy beam from the 60Co unit used at INOR was modeled using the EGS Monte Carlo radiation transport system. The rectangular geometry of the collimating system is fully simulated and the effect of considering the outer collimator a set of stepped collimators rather than a collimator with a sm...

Full description

Saved in:
Bibliographic Details
Main Authors: J. Ernesto Mainegra, Roberto Capote
Format: Article
Language:English
Published: Instituto Tecnológico de Costa Rica 2016-11-01
Series:Tecnología en Marcha
Online Access:https://revistas.tec.ac.cr/index.php/tec_marcha/article/view/2858
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The therapy beam from the 60Co unit used at INOR was modeled using the EGS Monte Carlo radiation transport system. The rectangular geometry of the collimating system is fully simulated and the effect of considering the outer collimator a set of stepped collimators rather than a collimator with a smooth surface was estimated. Variance reduction techniques are widely used to increase the efficiency of our calculations. Even for the dose at very shallow depths (<0.5 mm) the agreement of our calculations with the experimental results is very good. The shift of peak dose to about 1 mm due to electron contamination is observed.
ISSN:0379-3982
2215-3241