Usage of the micropixel detector TimePix for observation of the dynamics of phase transitions in metals

The results for the test of a TimePix hybrid detector as a tool for measuring and imaging in real time dynamics of phase transitions in metal alloys under heating or cooling are presented. The X-rays ( ~ 10 keV) experimental setup explored for the diffraction studies at the Institute for Problems of...

Full description

Saved in:
Bibliographic Details
Main Authors: M. V. Pugach, V. V. Burdin, M. Campbell, O. S. Kovalchuk, X. Llopart, M. V. Minakov, V. M. Militsiya, O. Yu. Okhrimenko, Yu. V. Podrezov, S. Pospisil, V. M. Pugatch, D. I. Storozhyk, L. Tlustos, S. O. Firstov, A. V. Chaus
Format: Article
Language:English
Published: Institute for Nuclear Research, National Academy of Sciences of Ukraine 2012-12-01
Series:Ядерна фізика та енергетика
Subjects:
Online Access:http://jnpae.kinr.kiev.ua/13.4/Articles_PDF/jnpae-2012-13-0382-Pugach.pdf
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The results for the test of a TimePix hybrid detector as a tool for measuring and imaging in real time dynamics of phase transitions in metal alloys under heating or cooling are presented. The X-rays ( ~ 10 keV) experimental setup explored for the diffraction studies at the Institute for Problems of Material Science NASU (Kyiv) is briefly described. An evolution of the diffraction maxima position (~ 20 μm accuracy) of the scattered X-rays was ob-served with exposures from a few dozen to a few hundred milliseconds under heating or cooling of the samples of Armco iron at a rate of 100 to 250 C/s. Data for the phase transitions (alpha-, gamma-phases observed) during heating/cooling (20 - 1250 °C) using the X-ray diffraction were measured.
ISSN:1818-331X