Entanglement due to Delayed Removal of a Buddy Wire

A buddy wire is often used to aid in the delivery of balloons and stents when negotiating tortuous or calcified vessels. We present a planned two-stent mini-crush intervention complicated by entanglement of the buddy wire with the second stent and subsequent distortion of the stent within the guidin...

Full description

Saved in:
Bibliographic Details
Main Authors: Marc Vorpahl, Melchior Seyfarth, Klaus Tiroch
Format: Article
Language:English
Published: Wiley 2014-01-01
Series:Case Reports in Cardiology
Online Access:http://dx.doi.org/10.1155/2014/513737
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A buddy wire is often used to aid in the delivery of balloons and stents when negotiating tortuous or calcified vessels. We present a planned two-stent mini-crush intervention complicated by entanglement of the buddy wire with the second stent and subsequent distortion of the stent within the guiding catheter. Based on this case, we suggest removing the buddy wire immediately after successful positioning of the first stent, because entrapment with a second stent is possible and may lead to challenging situations in a simultaneous two-stent strategy.