Investigation of the Acoustic Properties of Viscosity Standards

Longitudinal and shear ultrasonic wave velocities were measured versus temperature in the viscosity standards of Paragon S8000S, N30000S and Cannon N2700000. The measurements were performed by the through-transmission method at the frequency of 2 MHz. Ultrasonic pulses were sent via polymethyl metha...

Full description

Saved in:
Bibliographic Details
Main Authors: Rymantas KAZYS, Algirdas VOLEISIS, Reimondas SLITERIS
Format: Article
Language:English
Published: Institute of Fundamental Technological Research Polish Academy of Sciences 2015-10-01
Series:Archives of Acoustics
Subjects:
Online Access:https://acoustics.ippt.pan.pl/index.php/aa/article/view/1601
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Longitudinal and shear ultrasonic wave velocities were measured versus temperature in the viscosity standards of Paragon S8000S, N30000S and Cannon N2700000. The measurements were performed by the through-transmission method at the frequency of 2 MHz. Ultrasonic pulses were sent via polymethyl methacrylate (PMMA) waveguides between the tips of which a small amount of the particular standard liquid was placed. The velocities of longitudinal and shear waves were determined to depend on the viscosity of the liquid and increase with the viscosity.
ISSN:0137-5075
2300-262X